A&D SW-15KS User Manual

Page of 46
31
 
13.2.3. Command mode 
In the command mode, the scale is controlled by commands that come from an 
external device such as a personal computer. 
 
Command List
   
Command Description 
Remarks 
Send data immediately. 
 
Zero the scale when the weight is stable. 
Same as the    ZERO    switch.
Tare the scale when the data is stable. 
Same as the    TARE    switch. 
Switch the weighing unit. 
Same as the    UNITS    switch.
?H2 
When the five-level comparator mode is used: 
            Send the current HIHI limit value. 
When the three-level comparator mode is used: 
      Send the current HI limit value. 
?H1 
When the five-level comparator mode is used: 
      Send the current HI limit value. 
When the three-level comparator mode is used: 
            Not used 
?L1 
When the five-level comparator mode is used: 
      Send the current LO limit value. 
When the three-level comparator mode is used: 
            Not used 
?L2 
When the five-level comparator mode is used: 
            Send the current LOLO limit value. 
When the three-level comparator mode is used: 
      Send the current LO limit value. 
Send a setting value. 
 
Function settings 
Five-level (
Cp-l 0
Three-level (
Cp-l 1
H2 
When the five-level comparator mode is used: 
      Set the HIHI limit value. 
When the three-level comparator mode is used: 
      Set the HI limit value. 
H1 
When the five-level comparator mode is used: 
      Set the HI limit value. 
When the three-level comparator mode is used: 
            Not used. 
L1 
When the five-level comparator mode is used: 
      Set the LO limit value. 
When the three-level comparator mode is used: 
            Not used. 
L2 
When the five-level comparator mode is used: 
      Set the LOLO limit value. 
When the three-level comparator mode is used: 
      Set the LO limit value. 
Set the six-digit value excluding 
the polarity and decimal point