22
 
12. FUNCTIONS 
 
The scale has function settings to specify the scale performance. 
The parameters set in the function settings are maintained even if the power is turned 
OFF. 
 
 
 
 
 
12.1. Setting the parameters 
 
1. Turn the power OFF. 
2. Press and hold the    TARE    switch and press the 
 ON/OFF   switch to turn the power ON. The software 
version is displayed. 
‰
 “
∗∗∗” indicates the software version number. 
3. After about one second, the item is displayed. 
 
4. Change the item or parameter using the following switches. 
  UNITS        To display the next item. 
 PRINT    To increase the value of the blinking 
digit by one (to change the parameter). 
‰
 When the parameter is changed, the 
STABLE indicator turns off. 
 
5. Press the  TARE  switch to store the setting value.  
After “
end
”, the power is turned OFF automatically. 
 
‰
 To cancel the setting procedure without storing the value, press 
the    ZERO    switch. The power is turned OFF automatically. 
 
12.2. To restore the function settings to the 
factory setting values 
1. Turn the power OFF. 
2. Press and hold the    TARE    switch and press the 
  ON/OFF    switch to turn the power ON. Only release the 
 ON/OFF  switch and hold the  TARE  switch until 
Clrf no
” with “
no
” blinking is displayed. 
3. Press the  PRINT  switch. 
Clrf no
” changes to “
Clrf go
” with “
go
” blinking. 
4. When  “
Clrf go
” is displayed, press the  TARE  switch. 
The factory setting values are restored. After “
end
”, the power 
is turned OFF automatically. 
‰
 To cancel the restoring procedure, press the  ZERO  switch. 
The power is turned OFF automatically.. 
Item 
Parameter
While  
holding down,    press
Press
Press
When changed
Press
Press