A&D SW-6KS User Manual

Page of 46
23
 
12.3. Function list   
 
Item 
Parameter 
Description 
Š
  0 
Auto power-off disabled 
Auto power-off function 
poff 
 
Auto power-off enabled   
Turns the power OFF 
automatically  
 
Normal (1/3,000) 
Š
  1 
High (1/6,000 or 1/7,500) 
Weight display 
resolution 
re5o
 
 
Maximum (1/12,000 or 1/15,0 00)
 
Changes the   
minimum display 
Š
  0 
kg 
Weighing unit 
Ut-g
 
 
 
 
Zero tracking function disabled 
Zero tracking 
trc
 
Š 
Zero tracking function enabled 
Tracks the zero drift 
Š
  0 
±0.5d (width 1d) 
 
±1.0d (width 2d) 
Stability band 
width 
5t-b
 
 
±2.0d (width 4d) 
 
0.5 second 
Š
  1 
1.0 second 
Stability 
detection time 
5t-t
 
 
2.0 seconds 
Conditions to turn the 
STABLE indicator on
 
d = minimum display 
in kg
 
 
Weak stability / fast response 
Š
  1 
Normal stability / normal response 
Stability 
Response speed 
Cond
 
 
Strong stability / slow response 
Filtering 
Š
  0 
2400 bps 
 
4800 bps 
Serial interface 
baud rate 
bp5
 
 
9600 bps 
Š
  0 
7 bits / EVEN 
 
7 bits / ODD 
Serial interface 
data bits / parity 
btpr
 
 
8 bits / Non parity 
Š
  0 
Stream mode / Command mode 
 
Command mode only 
  2 
Print switch mode / 
Command mode
 
  3 
Auto-print mode +/- data / Command mode 
Serial interface 
Data output 
mode 
prt
 
  4 
Auto-print mode + data/ Command mode 
Š
  0 
RS-232C 
 
RS-422 
Serial interface 
Type 
5if
 
 
RS-485 
Serial interface 
Address 
adr
 
## 
## = 00 to 99 
(Factory setting: ## = 01) 
 
No reply except data to commands 
Serial interface 
Response 
aCk
 
Š
  1 
Reply to commands 
Only when the 
options, RS-232C or 
RS-422/485, are 
used. 
 
Š
  0 
Five-level (LOLO, LO, OK, HI, HIHI) 
Comparator mode 
Cp-l 
 
Three-level (LO, OK, HI)