AD-4405 Weighing Indicator
 
Weighing a Mass less than the Maximum Capacity 
Step12  Set a mass value using the 
0
 - 
9
 keys. 
Step13  Place a mass equivalent to the displayed value 
on the weighing unit. Proceed to step 15. 
 
Weighing a Maximum Capacity Mass 
Step14  Place a mass equivalent to the maximum 
capacity on the weighing unit. 
Step15 Press the 
ENTER
 key after the STABLE 
mark turns on. Proceed to step17. 
 
ESC
 key .......... The key to proceed to step17 
without changing the span 
parameters. 
Caution: Do not press the 
ENTER
 key while the 
STABLE mark is off (detecting motion). 
Arrange the condition using the F00 filter 
function. 
 
Digital Input 
Step16 Pressing 
the 
+/-
 key, a stored input voltage 
parameter of the span is displayed in the 
unit of mV/V. Adjust the input voltage using 
the 
0
 - 
9
 keys. (It is possible to store a 
greater value than the capacity.) 
 
ENTER
 key...... The key to store the span 
parameters and proceed to 
step 17. 
 
ESC
 key .......... The key to proceed to step 17 
without changing the span 
parameters. 
 
5.2.5. 
Exiting the Calibration Mode 
Step17 Check 
the 
Calend display. 
 
  Use the following keys. 
 
CAL
 key ..... To store the parameters and display   
Caloff. Proceed to the next step. 
 
 
Caloff 
could be displayed by pressing 
the
 
+/-
 key while pressing the 
ON/OFF
 key 
instead of 
CAL
 key.
 
5.2. Calibration Procedure 
Turn on STABLE mark
Span calibration 
Maximum capacity 
To step 14, step 16 
To step 12
New mass value 
Mass of 
maximum capacity 
To step 14 
To step 16 
Place the displayed
mass on the 
weighing unit 
step 13
step 14
Stored input voltage 
New input voltage 
mV/V 
mV/V 
Turn display off 
9
0
9
+/
+/
+/
+/-
--
-
NET/GROSS
PRINT
ENTER
PRINT
ENTER
PRINT 
ENTER
+/
+/
+/
+/-
--
-
NET/GROSS
PRINT
ENTER
or