5.2.3. Zero 
Calibration 
Procedure 
Step 8  Check the 
 Cal 0 
 display.  
Select a zero calibration method to adjust the zero point 
Weighing input 
(Normal way) 
The adjustment method 
with nothing on the 
weighing unit. 
To step 9 
Digital input 
The numerical way to 
enter a load cell output 
voltage. 
To step 10
 
Weighing Input 
Step 9  Place nothing on the weighing unit. Press the 
ENTER
 
key after the STABLE mark has turned ON. The new 
zero point parameter will be stored. Proceed to step11. 
 
ESC
 key .......... The key not to change the zero point 
data and proceed to the next step. 
Caution: Do not press the 
 ENTER 
  key while the STABLE 
mark is off (detecting motion). Arrange the 
condition using the F00 filter function. 
 
Digital Input 
Step10 Pressing 
the 
+/-
 key, a stored input voltage parameter 
of the zero point is displayed in the unit of mV/V. Adjust 
the input voltage using the 
0
 - 
9
 keys. 
 
ENTER
....... The key to store the zero point parameter 
and proceed to the next step. 
 
ESC k
ey..... The key to proceed to the next step without 
changing the parameter. 
 
 
 
5.2.4. Span 
Calibration 
Step11  Check the capacity display after 
Cal f 
  is displayed for 2 seconds. 
 
  Select a span calibration method to adjust the capacity. 
Weighing a mass less than the 
maximum capacity 
The method to weigh a mass less 
than the maximum capacity. 
To step 12
Weighing a maximum capacity 
mass 
The method to weigh a mass 
equivalent to the maximum capacity. 
To step 14
Digital input 
The numerical way to enter a load 
cell output voltage. 
To step 16
 
 
5.2. Calibration Procedure
Zero calibration 
Place nothing on 
the weighing unit
(No load) 
Turn on STABLE mark
Stored input voltage
New input voltage
mV/V
mV/V
To Span Calibration 
+/
+/
+/
+/-
--
-
NET/GROSS 
PRINT 
ENTER
PRINT
ENTER
0