AD-4405 Weighing Indicator
 
 
12.
12.
12.
12.    Counting Func
Counting Func
Counting Func
Counting Function
tion
tion
tion    
 
  This function determines the number of objects in a sample based on the unit 
weight. The unit weight is stored in one of the code memory data in non-volatile 
memory, and is not lost even if the AC power supply is disconnected. 
 
12.1.
12.1.
12.1.
12.1.    Using 
Using 
Using 
Using the 
the 
the 
the Counting Fu
Counting Fu
Counting Fu
Counting Function
nction
nction
nction    
 
 Preparation: 
Set 
f07 1 in the F-Functions. 
 
  Select a code memory number. Refer to “9.1.1. Using Code Memory”. 
 
  Register the unit weigh. Refer to “12.2. Unit Weight Registration”. 
 
 Press 
the 
MODE
 key to enter the counting mode. The unit in the display changes to 
pcs
pcs
pcs
pcs. 
 
  In the counting mode, TARE, ZERO, NET/GROSS and other operations are the same 
as that in the normal weighing mode. 
 
12.2.
12.2.
12.2.
12.2.    
Unit Weight Registration 
Unit Weight Registration 
Unit Weight Registration 
Unit Weight Registration      
 
  There are two methods to register the unit weight. One is to weigh the actual samples 
for unknown object's unit weight and the other way is by digital input of the unit weight 
by key or via the serial interface. 
 
  The unit weight is one item of the code memory data. Select the code memory 
number and start the unit weight registration in the weight display or piece count 
display. The unit weight registration can not be done from the code memory 
number display. 
 
  There is no counting accuracy improvement function by re-calculating the unit 
weight based on the actual weight. 
 
Weigh Actual Samples 
 
When the unit weight is unknown, prepare some samples to determine the unit weight. 
Step  1  Enter the counting mode. 
Step 2  Press the 
SAMPLE
 key to enter the unit weight registration.   5  0  or  5  -  is 
displayed and unit 
pcs
pcs
pcs
pcs blinks. The left side of the display is the number of samples. 
The right side of the display indicates the weight, 0   means that the weight is zero 
and   -   means that the display is not at zero. 
Step  3  Select the number of samples and place the samples specified on the weighing 
sensor unit. Using following keys: 
 
0
 - 
9
 key..................The key to set the number of samples. Maximum is 9999.   
 
+/-
 key .....................The key to select the number of samples, 5, 10, 20, 50,100. 
 
ZERO
 key................The key to set the zero point. 
 
TARE
 key ................The key to tare. When using a container. 
12. Counting Function