AD-4405 Weighing Indicator
 
 
11.
11.
11.
11.    Hold Function
Hold Function
Hold Function
Hold Function    
 
  This function displays the hold weight data after averaging the weight data for a 
specific period. 
 
  Useful to determine a living animal’s weight. 
 
  Averaging time can be set up to 9.9 seconds by a 0.1 second step. 
 
  3 methods are available to start averaging; manual start, start automatically after 
stable and manual / automatic start. 
 
  Manual start is available with key switch or external input. 
 
  Serial interface commands are also available; averaging start, releasing the hold 
data and outputting the hold state. Refer “16.3.4. Commands for Hold Function”. 
Caution 
  This function can not be use under the setting with 
Cf00 
Cf00 
Cf00 
Cf00 1111. 
 
  Averaging can not start at a displayed value smaller than +/- 5 digits. 
 
  Data when the display is over is not included for averaging. 
 
  When powered off, the hold is released automatically. 
 
  There is no peak hold function. 
 
The Display and the Data Output of Hold and Average
The Display and the Data Output of Hold and Average
The Display and the Data Output of Hold and Average
The Display and the Data Output of Hold and Average    
 
  The weight display is blinking during the averaging period. 
 
  The output data in the averaging period is the actual weight at the time. 
 
  The weighing unit is blinking when the weight display is in the hold state. 
 
  The output data format of the hold weight data is the same as that of the normal 
weight data except the header of stable state is “HD” in the response to the “RW” 
or “RW,n”(n=1 or 2) command. 
 
Relations to the Other Functions
Relations to the Other Functions
Relations to the Other Functions
Relations to the Other Functions         
 
  If automatic accumulation (f20 21) and/or auto print (f37 2, 3 f40 2, 3) is set, 
accumulation and/or data output is performed after determining the hold data. 
 
11.1.1.  Setting the Hold Functions   
 
  F27 determines the method of starting the average. 
f27 1 
Manual start: Start the average and release with key switch 
operation. 
f27 2 
Automatic start: After passing the inhibit region * and detect stable **, 
starts the average automatically, releasing the data when the weight 
returns to inhibit region. 
f27 3 
Both Manual start and Automatic start. 
  *Inhibit region 
0 +/- 5digit 
**Stable detection 
Satisfied both f02 and f03 
11. Hold Function