7. CALIBRATION 
 
This function adjusts the balance for accurate weighing. 
Perform a calibration in the following cases. 
 
‰
 When the balance is first used. 
‰
 When the balance has been moved. 
‰
 For regular calibration.
 
 
 
 
 
 
 
 
 
 
7-1. Calibration using a weight 
‰
 Prepare the calibration weight (sold separately) in advance.
 
 
1. Warm up the balance for at least half an hour with 
nothing on the pan. 
 
2. Press and hold the calibration (CAL) switch until   
  Cal          appears, and release the switch. 
 
3. The balance displays   Cal 0  .  
 
To change the calibration weight value, proceed to 
step 4. 
 
To use the calibration weight value in the balance 
memory, proceed to step 5. 
 
4. Press the  SAMPLE  key. The display shows the 
calibration weight value in “gram” that is stored in the 
balance. Use the following keys to change the value. 
 
 
 SAMPLE  key  To select the digit blinking to 
change. 
 
 RE-ZERO  key  To set the value of the digit 
selected. 
 
 PRINT  key 
To store the value and return to 
step 3. 
 
  UNITS    key 
To cancel the value and return to 
step 3. 
 
 
 
 
  Cal 
  Cal 0 
 06000.0 
  Cal 0 
Set the weight using
the relevant keys. 
Release the CAL switch.
 
 
Calibration (CAL) switch 
Calibration (CAL) switch cover
Press and hold
the CAL switch.