A&D EJ-1500 User Manual

Page of 42
 16
7-2. Gravity acceleration correction   
 
When the balance is first used or has been moved to a different place, it should be 
calibrated using a calibration weight. 
But if the calibration weight cannot be prepared, the gravity acceleration correction will 
compensate the balance. Change the gravity acceleration value of the balance to the 
value of the area where the balance will be used. See the gravity acceleration map 
appended to the end of this manual. 
 
‰
 Note 
  Gravity acceleration correction is not required when the balance is 
calibrated using a calibration weight at the place where the balance 
is to be used.   
 
 
1. Press and hold the calibration (CAL) switch until   
  Cal          appears, and release the switch. 
 
2. The balance displays   Cal 0  . 
 
 
3. Press the    RE-ZERO    key. 
The display shows the gravity acceleration value 
stored in the balance. 
Use the following keys to change the value. 
 
 SAMPLE  key  To select the digit blinking to 
change. 
 
 RE-ZERO  key  To set the value of the digit 
selected. 
 
 PRINT  key 
To store the value and return to 
step 2. 
 
 UNITS   key 
To cancel the value and return to 
step 2. 
 
4. After setting the value, press the    PRINT    key. 
 Cal 0    is displayed again. 
 
5. If it is necessary to calibrate the balance using a 
calibration weight, go to step 4 of 7-1. 
To finish the setting, press    UNITS    key. 
 
6.  end     appears and the balance returns to the 
weighing mode. 
 
 
 
 
 
 
 
 
  Cal 
  Cal 0 
  9.7985 
 Cal 0 
Returns to   
the weighing mode
To store
Release the CAL switch.
Set the value using
the relevant keys. 
Press and hold
the CAL switch.