Tektronix 8.00E+04 User Manual

Page of 68
Specifications
48
80E00 Electrical Sampling Modules User Manual
Table 7: Electrical sampling modules - Signal acquisition
Specifications
Characteristics
Real time accessory
interface
Tekprobe--SMA interface is provided through the electrical
sampling-module interface, one per vertical channel.
Channel input
Sampling module
Input connector
p
connector
80E02, 80E03, 80E04
Precision 3.5 mm female
connector.
80E01
Precision 2.4 mm female
connector (2.4 mm male to
2.92 mm (K) female adapter,
015-0703-xx, is supplied).
80E06
Precision 1.85 mm female
connector (V) (2.4 mm male to
2.92 mm (K) female adapter,
015-0703-xx, is supplied).
Number of input
Sampling module
Channels
p
channels
80E01, 80E06
1
80E02, 80E03, 80E04:
2
n Input impedance
50
Ω ±0.5 Ω
Vertical dynamic
range
1 V
pp
(offset
±500 mV)
Vertical operating
range
1
, maximum
±1.6 V
Vertical nondestruct
2
Sampling module
Maximum input
range
2
(Maximum input
80E01, 80E06
2.0 V (DC+peak AC)
(Maximum input
voltage)
80E02, 80E03, 80E04
3.0 V (DC+peak AC)
Vertical number of
digitized bits
14 bits full scale
Vertical sensitivity
range
3
The range of available full scale input settings.
10 mV to 1 V full scale
Compensation
temperature range
±5_ C about temperature where compensation was performed. If
compartment is changed on the mainframe, a sampling module
extender is employed, or the length of the sampling module extender is
changed, the channel must be recompensated.
n DC voltage
accuracy, single point,
within
5_C of
compensated
temperature
±2 mV ±0.007 (assigned offset)
±0.02 (vertical value -- assigned offset)
n DC vertical
voltage deviation from
linear least squares fit
±10 mV