Tektronix 8.00E+04 User Manual

Page of 68
Specifications
80E00 Electrical Sampling Modules User Manual
49
Table 7: Electrical sampling modules - Signal acquisition (cont.)
Specifications
Characteristics
n Rise time
4
Sampling module
Rise time
80E01
≤ 7 ps, typical
80E02
≤ 28 ps
80E03 and 80E04
≤ 17.5 ps
80E06
≤ 5.0 ps, typical
n Analog bandwidth
5
Sampling module
Bandwidth
80E01
50 GHz
80E02
12.5 GHz, typical
80E03 and 80E04
20 GHz, typical
80E06
65 GHz
70 GHz, typical
Step response
7
Sampling module
Aberrations, step transition
6
p
p
aberrations
7
, typical
80E02, 80E03 and 80E04
±3% or less over the zone 10 ns
to 20 ps before step transition
+10%, --5% or less for the first
300 ps following step transition
±3% or less over the zone
300 ps to 5 ns following step
transition
±1% or less over the zone 5 ns
to 100 ns following step transition
±0.5% after 100 ns following
step transition
80E01
±3% or less over the zone 10 ns
to 20 ps before step transition
+12%, --5% or less for the first
300 ps following step transition
+5.5%, --3% or less over the zone
300 ps to 3 ns following step
transition
±1% or less over the zone 3 ns
to 100 ns following step transition
±0.5% after 100 ns following
step transition
Step response
overshoot
7
, typical
80E06
+ 5% or less for the first 300 ps
following step transition