General Troubleshooting & Repair of the M703E Calibrator 
M703E Calibrator Operators Manual 
PHOTOMETER PRESSURE SENSOR:  
1.  Measure the pressure on the inlet side of S1 with an external pressure meter. 
2.  Measure the voltage across TP4 and TP1.   
  The expected value for this signal should be: 
 
EXAMPLE:  If the measured pressure is 20 Hg-in-A, the expected voltage level between TP4 and 
TP1 would be between 2870 mVDC and 3510 mVDC. 
EXAMPLE:  If the measured pressure is 25 Hg-in-A, the expected voltage level between TP4 and 
TP1 would be between 3533 mVDC and 4318 mVDC. 
  If this voltage is out of range, then either pressure transducer S1 is bad, the board is bad or there is a 
pneumatic failure preventing the pressure transducer from sensing the absorption cell pressure 
properly. 
 
O
3
 GENERATOR PRESSURE SENSOR   
1.  Measure the pressure on the inlet side of S2 with an external pressure meter. 
2.  Measure the voltage across TP5 and TP1.   
  The expected value for this signal should be: 
   
EXAMPLE:  If the measured pressure is 25 psig, the expected voltage level between TP4 and TP1 
would be between 3470 mVDC and 4245 mVDC. 
EXAMPLE:  If the measured pressure is 30 psig, the expected voltage level between TP4 and TP1 
would be between 4030 mVDC and 4930 mVDC. 
  If this voltage is out of range, then either pressure transducer S1 is bad, the board is bad or there is a 
pneumatic failure preventing the pressure transducer from sensing the absorption cell pressure 
properly. 
 
PHOTOMETER FLOW SENSOR 
  Measure the voltage across TP3 and TP1.   
  With proper flow (800 cc
3
/min through the photometer), this should be approximately 4.5V (this 
voltage will vary with altitude).   
  With flow stopped (photometer inlet disconnected or pump turned OFF) the voltage should be 
approximately 1V.   
  If the voltage is incorrect, the flow sensor S3 is bad, the board is bad or there is a leak upstream of 
the sensor. 