General Troubleshooting & Repair of the M703E Calibrator 
M703E Calibrator Operators Manual 
For each of the steps the output should be within 1% of the nominal value listed in the table below except for the 
0% step, which should be within 0mV ±2 to 3 mV.  Make sure you take into account any offset that may have 
been programmed into channel (See Section 6.9.1.5).  
Table 11-9:  Analog Output Test Function - Nominal Values Voltage Outputs 
 
 
FULL SCALE OUTPUT OF  VOLTAGE RANGE 
(see Section 6.9.1.3) 
   100MV 1V 
5V 
10V 
STEP 
NOMINAL OUTPUT VOLTAGE 
1 0 
0  0  0 
2 20  20 
mV  0.2  1  2 
3 40  40 
mV  0.4  2  4 
4 60  60 
mV  0.6  3  6 
5 80  80 
mV  0.8  4  8 
6 100  100 
mV  1.0  5  10 
 
If one or more of the steps fails to be within these ranges, it is likely that there has been a failure of the either or 
both of the DACs and their associated circuitry on the motherboard. 
 
 
11.4.8.3. Status Outputs 
To test the status output electronics: 
1.  Connect a jumper between the “D“ pin and the “” pin on the status output connector. 
2.  Connect a 1000 ohm resistor between the “+” pin and the pin for the status output that is being tested. 
3.  Connect a voltmeter between the “” pin and the pin of the output being tested (see table below).   
4. Under 
the 
DIAG SIGNAL I/O menu (See Section11.1.3), scroll through the inputs and outputs until 
you get to the output in question.  
5.  Alternately, turn on and off the output noting the voltage on the voltmeter. 
   It should vary between 0 volts for ON and 5 volts for OFF.   
Table 11-10: 
Status Outputs Check 
PIN  
(LEFT TO RIGHT)
STATUS 
ST_SYSTEM_OK 
SPARE 
ST_CAL_ACTIVE 
ST_DIAG_MODE 
ST_TEMP_ALARM 
ST_PRESS_ALARM 
7 and 8 
SPARE 
192 
 
05744 Rev B