Working with the DIAGNOS Function 
Overview 
The DIAGNOS function displays sensor data and status information useful to a service technician for 
troubleshooting various fault conditions.  Explanations of the data shown in these screens are given below.  
Keypad Functions 
From the Data Display screen, use the Keypad buttons to place the arrow (>) on the display next to the 
DIAGNOS function. Then press ENTER to display the first of two Diagnostic screens.  
Press the Up Arrow Keypad button to toggle between the First and Second Diagnostic screen. 
First Diagnostic Screen 
 
 
PPM Level – Parts Per Million Level is the current detected gas level, and is the volume concentration 
referenced to standard temperature and pressure. 
Average Absorption Unit – This is the optical absorbency. In the absence of refrigerant the absorbency 
is 0.00000 au.  When sampling refrigerants, its value varies proportionally with the refrigerant 
concentration. 
µMoles/Liter – This is the absolute concentration in micro-moles per liter of refrigerant. 
Detector Voltage – This is a running average of the IR detector’s bench voltage as displayed in the 
second Diagnostic screen. 
Purge Valve Asterisk (– When viewing the diagnostic screen, the purge valve can be opened and 
closed by pressing the Left Arrow Keypad button. An asterisk appears on the display when the purge 
valve is open
Second Diagnostic Screen 
 
 
In the second diagnostic screen, toggle between displaying the Bench Voltage and the Noise value, and the 
Detector Temperature and Box Temperature by pressing the Right Arrow Keypad button. 
 
 
Bench Voltage – This is the current peak-to-peak output of the IR detector. In the absence of refrigerant 
this value can range from 3.90000V to 4.50000V. 
Noise –The Noise value is a 16 point running average of the noise portion of the IR detector’s output. This 
reading is valuable mainly when refrigerant is NOT present. 
Detector Temperature – This is the current detector temperature in °C. 
Box Temperature
 – This is the current internal enclosure temperature in °C. 
Fault Code – Current fault code. A value of <0000> indicates that no faults are being detected. 
Pressure Reading – This is the pressure as measured every purge cycle with the sample pump off and 
the gas-sample inlet open.  Its value is weather and altitude dependent and can range from 10.0 to 
15.5 PSIA. 
Purge Valve Asterisk (– When viewing the diagnostic screen, the purge valve can be opened and 
closed by pressing the Left Arrow Keypad button. An asterisk appears on the display when the purge 
valve is open
 
4.26500v  <0000> 
24.5cD* 14.00psi 
    0.4ppm  0.00 
0.00002au*4.260v 
Bench Voltage 
Fault Code  
Detector Temperature °C  
∗ Pressure Reading 
PPM Level 
µMole/Liter 
Avg. Absorption Unit     
∗ Detector 
Voltage 
0.00250n  <0000> 
29.5cB* 14.00psi 
Noise 
Fault Code  
Box Temperature °C     
∗ Pressure 
Reading