Bacharach 3015-4256 User Manual

Page of 36
22 
Instruction 3015-4256 
Working with the P-CHK Function 
Overview 
The P-CHK function (Pressure Check Function) (Page 17) is useful to a service technician for 
troubleshooting a flow fault problem. The monitor will trigger a flow fault if the pressure drop from 
ambient is less than 0.2 psi during a purge cycle, and 0.5 psi during a measurement cycle.  
Keypad Functions 
The Left Arrow Keypad button toggles the purge valve open and closed. Note that an asterisk (*) 
appears when the purge valve is open (purging).  
The Down Arrow Keypad button toggles the pump ON and OFF.  
Pressing the ENTER button stores the current manifold pressure shown on the left to the ambient 
pressure shown on the right (must be done with the pump OFF). 
Screen Display 
 
 
Current Manifold Pressure – Current manifold pressure in psia. 
Stored Ambient Pressure – Stored ambient pressure in psia. 
Pressure Difference – The difference between the current manifold pressure and the stored ambient 
pressure. 
Fault Code – Current fault code (Page 19). 
Purge Valve Asterisk (– When viewing the pressure-check screen, the purge valve can be opened 
and closed by pressing the Left Arrow Keypad button. An asterisk appears on the display when the 
purge valve is open
 
14.07 * AMB14.05 
-0.02dif  <1000> 
Current Manifold Pressure    
∗  Stored Ambient Pressure 
Pressure Difference 
 
Fault Code