Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
Calibration Procedures 
 
205 
9.6.  CALIBRATION OF THE M300E/EM’S ELECTRONIC 
SUBSYSTEMS 
9.6.1. DARK CALIBRATION TEST 
The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of the 
sync/demod board circuitry.  This allows the instrument to compensate for any voltage levels inherent in the 
sync/demod circuitry that might effect the calculation of CO concentration.   
Performing this calibration returns two offset voltages, one for CO MEAS and one for CO REF that are 
automatically added to the CPU’s calculation routine.  The two offset voltages from the last calibration procedure 
may be reviewed by the user via the front panel display. 
To activate the dark calibration procedure or review the results of a previous calibration, press: 
Offset for CO REF signal
SAMPLE
RANGE=50.0 PPM          CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X         
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X       
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG                        
SIGNAL I/O
PREV
NEXT
ENTR
EXIT
SETUP X.X     
ENTER PASSWORD
8
1
8
ENTR
EXIT
Continue pressing NEXT until ...
DIAG OPTIC            
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG DARK   
CO DARK CALIBRATION
VIEW
CAL
EXIT
DIAG DARK   
REF DARK OFFSET: 0.0mV
EXIT
Offset for CO MEAS signal
DIAG DARK   
MEAS DARK OFFSET: 0.0mV
EXIT
DIAG DARK   
DARK CAL 1% COMPLETE
EXIT
DIAG DARK   
DARK CALIBRATION ABORTED
EXIT
Calibration runs automatically
 