Calibration Procedures 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
206 
9.6.2. PRESSURE CALIBRATION 
A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas.  This data is 
used to compensate the final CO concentration calculation for changes in atmospheric pressure and is stored in 
the CPU’s memory as the test function PRES (also viewable via the front panel).   
NOTE 
This calibration must be performed when the pressure of the sample gas is equal to ambient 
atmospheric pressure.   
 Before performing the following pressure calibration procedure, disconnect the sample gas pump and 
the sample gas-line vent from the sample gas inlet on the instrument’s rear panel. 
To cause the analyzer to measure and record a value for PRES, press.    
SETUP X.X       
 CFG DAS RNGE PASS CLK
EXIT
SETUP X.X         
 COMM VARS
EXIT
SETUP X.X    
EXIT
ENTR
Continue pressing 
  until ...
 discards the 
new setting.
 accepts the 
new setting.
DIAG              
PREV NEXT
EXIT
DIAG    PCAL 
EXIT
DIAG    PCAL 
 
04288D DCN5752