Troubleshooting & Repair 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
300 
 
13.6.2. REMOVING/REPLACING THE GFC WHEEL  
When removing or replacing the GFC Wheel it is important to perform the disassembly in the following order to 
avoid damaging the components: 
1.  Turn off the analyzer. 
2.  Remove the top cover. 
3.  Open the instrument’s hinged front panel.   
4.  Locate the GFC Wheel/motor assembly.  See Figure 3-4. 
5.  Unplug the following electronic components: 
  The GFC Wheel housing temperature sensor 
 GFC Wheel heater 
  GFC Wheel motor power supply 
SAFETY SHIELD
SYNCHRONOUS MOTOR
SOURCE ASSEMBLY
THERMISTOR
HEATER
 
Figure 13-15:  Opening the GFC Wheel Housing 
 
6.  Remove the three (3) screws holding the opto-pickup printed circuit assembly to the GFC Wheel 
housing. 