Troubleshooting & Repair 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
302 
 
10.  Remove the one (1) screw fastening the GFC Wheel/mask assembly to the GFC motor hub. 
 
11
 
12
 
Figure 13-18:  Removing the GFC Wheel 
 
11.  Remove the GFC Wheel/mask assembly. 
12.  Follow the previous steps in reverse order to put the GFC Wheel/motor assembly back together. 
 
 
 
13.6.3. CHECKING AND ADJUSTING THE SYNC/DEMODULATOR,  CIRCUIT 
GAIN (CO MEAS) 
13.6.3.1. Checking the Sync/Demodulator Circuit Gain 
The M300E/EM Analyzers will operate accurately as long as the sync/demodulator circuit gain is properly 
adjusted.  To determine if this gain factor is correct: 
1.  Make sure that the analyzer is turned on and warmed up. 
2.  Set the analyzer display to show the STABIL or CO STB test function. 
3.  Apply Zero Air to Sample Inlet of the analyzer. 
4.  Wait until the stability reading falls below 1.0 ppm. 
5.  Change the analyzer display to show the CO MEAS 
  The value of CO MEAS must be > 2800 mV and < 4800 mV for the instrument to operate correctly. 
  Optimal value for CO MEAS is 4500 mV ± 300 mV.  If it is not, adjust the value. 