Troubleshooting & Repair 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
304 
 
13.6.4. DISK-ON-MODULE REPLACEMENT PROCEDURE 
Replacing the Disk-on-Module (DOM) will cause loss of all DAS data; it also may cause loss of some instrument 
configuration parameters unless the replacement DOM carries the exact same firmware version. Whenever 
changing the version of installed software, the memory must be reset. Failure to ensure that memory is reset can 
cause the analyzer to malfunction, and invalidate measurements. After the memory is reset, the A/D converter 
must be re-calibrated, and all information collected in Step 1 below must be re-entered before the instrument will 
function correctly. Also, zero and span calibration should be performed. 
1.  Document all analyzer parameters that may have been changed, such as range, auto-cal, analog 
output, serial port and other settings before replacing the DOM  
2.  Turn off power to the instrument, fold down the rear panel by loosening the mounting screws. 
3.  When looking at the electronic circuits from the back of the analyzer, locate the Disk-on-Module in the 
right most socket of the CPU board.  
4.  The DOM should carry a label with firmware revision, date and initials of the programmer.  
5.  Remove the nylon fastener that mounts the DOM over the CPU board, and lift the DOM off the CPU. Do 
not bend the connector pins. 
6.  Install the new Disk-on-Module, making sure the notch at the end of the chip matches the notch in the 
socket.  
7.  It may be necessary to straighten the pins somewhat to fit them into the socket. Press the DOM all the 
way in and reinsert the offset clip. 
8.  Close the rear panel and turn on power to the machine. 
9.  If the replacement DOM carries a firmware revision, re-enter all of the setup information.  
13.7. TECHNICAL ASSISTANCE 
If this manual and its troubleshooting / repair sections do not solve your problems, technical assistance may be 
obtained from: 
Teledyne API, Customer Service, 
9480 Carroll Park Drive  
San Diego, California 92121-5201USA 
Toll-free Phone: 800-324-5190 
Phone: 858-657-9800 
Fax: 858-657-9816 
Email: api-customerservice@teledyne.com 
Website: http://www.teledyne-api.com/ 
Before you contact Teledyne API Customer service, fill out the problem report form in Appendix C, which is also 
available online for electronic submission at 
 
 
04288D DCN5752