Teledyne M300EM User Manual

Page of 398
Getting Started 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
50 
3.5.4. FUNCTIONAL CHECK 
After the analyzer’s components have warmed up for at least 60 minutes, verify that the software properly 
supports any hardware options that were installed.   
For information on navigating through the analyzer’s software menus, see the menu trees described in  
Appendix A.1.   
  Check to make sure that the analyzer is functioning within allowable operating parameters.   
  Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as their 
expected values.   
  These functions are also useful tools for diagnosing performance problems with your analyzer (see 
  The enclosed Final Test and Validation Data Sheet (P/N 04307) lists these values before the 
instrument left the factory. 
To view the current values of these parameters press the following key sequence on the analyzer’s front 
panel.  Remember that until the unit has completed its warm-up these parameters may not have stabilized.   
 
If your analyzer has an Ethernet card installed and your network is running a Dynamic Host Configuration 
Protocol (DHCP) software package, the Ethernet option will automatically configure its interface with your LAN.   
  However, it is a good idea to check these settings to make sure that the DHCP has successfully 
downloaded the appropriate network settings from your network server (See Section 8.4.2). 
  If your network is not running DHCP, you will have to configure the analyzer’s interface manually (See 
04288D DCN5752