Teledyne M300EM User Manual

Page of 398
Getting Started 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
52 
3.6.2.1. Verifying the M300E/EM Reporting Range Settings 
While it is possible to perform the following procedure with any range setting we recommend that you perform 
this initial checkout using following reporting range settings: 
  Unit of Measure: PPM 
  Analog Output Reporting Range: 50 ppm  
  Mode Setting: SNGL 
While these are the default setting for the M300E/EM Analyzer, it is recommended that you verify them before 
proceeding with the calibration procedure, by pressing: 
 SETUP X.X      
RANGE MODE:SINGL  
SNGL DUAL AUTO                         
ENTR
EXIT
SAMPLE             RANGE=50.0 PPM        CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X         
PRIMARY SETUP MENU
 CFG DAS
RNGE
PASS CLK MORE
EXIT
 SETUP X.X      
RANGE CONTROL MENU
  MODE
SET UNIT
DIL
EXIT
 SETUP X.X 
 RANGE: 50.0 Conc
0
0
0
5
0
.0
ENTR
EXIT
 SETUP X.X      
CONC UNITS:PPM
PPB PPM UGM MGM
ENTR
EXIT
 SETUP X.X      
RANGE CONTROL MENU
  MODE
SET
UNIT
DIL
EXIT
 SETUP X.X      
RANGE CONTROL MENU
  MODE
SET UNIT
DIL
EXIT
Verify that the MODE
is set for SNGL.
If it is not, press
SINGL  ENTR.
Press EXIT
3x’s to return 
the M200E to 
the SAMPLE 
mode.
Verify that the RANGE is 
set for 50.0
If it is not, toggle each 
numeric key until the 
proper range is set, then 
press ENTR.
Verify that the UNIT
is set for PPM
If it is not, press
PPM  ENTR.
 
04288D DCN5752