3.6.2.1. Verifying the M300E/EM Reporting Range Settings
While it is possible to perform the following procedure with any range setting we recommend that you perform
this initial checkout using following reporting range settings:
this initial checkout using following reporting range settings:
Unit of Measure: PPM
Analog Output Reporting Range: 50 ppm
Mode Setting: SNGL
Analog Output Reporting Range: 50 ppm
Mode Setting: SNGL
While these are the default setting for the M300E/EM Analyzer, it is recommended that you verify them before
proceeding with the calibration procedure, by pressing:
proceeding with the calibration procedure, by pressing:
SETUP X.X
RANGE MODE:SINGL
SNGL DUAL AUTO
ENTR
EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
DIL
EXIT
SETUP X.X
RANGE: 50.0 Conc
0
0
0
5
0
.0
ENTR
EXIT
SETUP X.X
CONC UNITS:PPM
PPB PPM UGM MGM
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
DIL
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
DIL
EXIT
Verify that the MODE
is set for SNGL.
If it is not, press
SINGL ENTR.
Press EXIT
3x’s to return
the M200E to
the SAMPLE
mode.
Verify that the RANGE is
set for 50.0
If it is not, toggle each
numeric key until the
proper range is set, then
press ENTR.
Verify that the UNITs
is set for PPM
If it is not, press
PPM ENTR.
04288D DCN5752