Basic Operation 
 
85 
6. 
BASIC OPERATION  
The M300E/EM Analyzer is a computer-controlled analyzer with a dynamic menu interface that allows all major 
operations to be controlled from the front panel display and keyboard through user-friendly menus (a complete 
set of menu trees is located in Appendix A of this manual). 
This section includes step-by-step instructions for using the display/keyboard to set up and operate the 
M300E/EM Analyzer's basic CO measurement features and functional modes.   
6.1. OVERVIEW OF OPERATING MODES 
The M300E/EM software has a variety of operating modes.  Most commonly, the analyzer will be operating in 
Sample Mode.  In this mode a continuous read-out of the CO concentration is displayed on the front panel.  If 
the analyzer is configured to measure more than one gas (e.g. CO along with O
2
 or CO
2
) the display will cycle 
through gas list. 
While in SAMPLE mode calibrations can be performed and TEST functions as well as WARNING messages can 
be examined.  If any of the analyzer’s analog outputs are enabled, the current concentration value will be 
available at the analog output connector. 
The second most important operating mode is SETUP mode.  This mode is used for performing certain 
configuration operations, such as programming the iDAS system or the configurable analog output channels, or 
setting up the analyzer’s serial communication channels (RS-232/RS-485/Ethernet).  The SETUP mode is also 
used for performing various diagnostic tests during troubleshooting. 
 
Mode Field 
SAMPLE               RANGE=50.00 PPM          CO= XX.XX
<TST    TST>  CAL                                             SETUP
Concentration  Field
Message  Field 
 
Figure 6-1: 
Front Panel Display 
The mode field of the front panel display indicates to the user which operating mode the unit is currently running.   
Besides SAMPLE and SETUP, other modes the analyzer can be operated in are: 
Table 6-1: 
Analyzer Operating Modes 
MODE EXPLANATION 
SAMPLE 
Sampling normally, flashing text indicates adaptive filter is on. 
SAMPLE A
 
Indicates that unit is in Sample Mode while AUTOCAL feature is active (Internal Span Only). 
M-P CAL 
This is the basic calibration mode of the instrument and is activated by pressing the CAL key. 
SETUP [X.X] 
SETUP mode is being used to configure the analyzer.  The gas measurement will continue during this 
process.  The revision of the M300E/EM firmware being run will appear after the word “SETUP” 
ZERO CAL [type] 
1,2 & 3
 
Unit is performing ZERO calibration procedure. 
LO CAL [type]
 2 & 3
 
Unit is performing LOW SPAN (midpoint) cal check procedure. 
SPAN CAL [type]
 1,2 & 3
 
Unit is performing SPAN calibration procedure. 
DIAG Mode 
One of the analyzer’s diagnostic modes is active (Section 7.3).   
[type:] 
1
A: Initiated automatically by the AUTOCAL feature (Internal Span Only). 
2
M: initiated manually by the user via the front panel controls. 
3
R:  initiated remotely through the COM ports or digital control inputs. 