Teledyne M300EM User Manual

Page of 398
Basic Operation 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
86 
6.2. SAMPLE MODE 
This is the analyzer’s standard operating mode.  In this mode the instrument is analyzing the gas in the sample 
chamber, calculating CO concentration and reporting this information to the user via the front panel display, the 
analog outputs and, if set up properly, the RS-232/RS-485/Ethernet ports.   
 
NOTE 
A value of “XXXX” displayed in the CO Concentration field means that the M/R ratio is invalid because 
CO REF is either too high (> 4950 mVDC) or too low (< 1250 VDC). 
 
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range reading or the 
analyzer’s inability to calculate it. 
A variety of TEST functions are available for viewing at the front panel whenever the analyzer is at the MAIN 
MENU
.  These functions provide information about the various functional parameters related to the analyzers 
operation and its measurement of gas concentrations.  This information is particularly a performance problem 
during troubleshooting (see Section 13.1.2).  
To view these TEST functions, press, 
 
Figure 6-2: 
Viewing M300E/EM Test Functions 
 
 
NOTE 
All pressure measurements are represented in terms of absolute pressure. Absolute, atmospheric 
pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300 m gain in altitude. A variety of 
factors such as air conditioning and passing storms can cause changes in the absolute atmospheric 
pressure. 