Maintenance and Service
Section 5: Maintenance and Service
6888Xi Advanced Electronics Instruction Manual
December 2012
PN 51-6888Xi
FIGURE 5-11. DR Board Wiring, Left-Hand Side
ii.
Press the <EXIT> button once to enteer the “Diagnostic Menu”.
iii.
Use the Up/Down keys to select “3-Acknowledge Alarms”. Press <ENTER>.
iv.
Use the Up/Down keys to select “3-Ack Probe Changed”. Press <ENTER>.
v.
Use the Up/Down keys to select “1-Get From Probe”. Press <ENTER> twice.
vi.
When transfer is complete, press <EXIT> twice to return to the main menu.