Maintenance and Service
91
6888Xi Advanced Electronics Instruction Manual
Section 5: Maintenance and Service
PN 51-6888Xi
December 2012
FIGURE 5-10. DR Board Wiring, Right-Hand Side
12. Reapply power to the Remote Interface and Probe. When initialization is complete, the
Probe will begin heating up, but an alarm condition will be shown. Prior to using the
Transmitter for measurement purposes, perform one of the following:
a.
Recalibrate the O
2
Probe according to the applicable calibration instructions in Section 3.
b.
Synchronize the calibration parameters as follows:
i. 
Press the <DIAG> button once to view the alarms. A “Probe Changed” alarm will be
shown.