Oritron ICS-2100 User Manual
ICS-2100 Ion Chromatography System
90
Doc. 065291-01 3/09
4.5
Pump Does Not Start
•
No power (front
Power
LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace, if needed (see
•
No communication between ICS-2100 and Chromeleon
Verify that the USB cable is connected correctly. For connection and setup
information, refer to the ICS-2100 installation instructions.
information, refer to the ICS-2100 installation instructions.
•
Pump is turned off
Turn on the pump from either the Chromeleon Control panel or the ICS-2100
touch screen
touch screen
HOME
page.
•
Flow rate is set to 0
Set the flow rate from either the Chromeleon Control panel or the ICS-2100
touch screen
touch screen
HOME
page.
4.6
No Flow
•
Pump waste or priming valve open (see
Close the valves by turning the knobs clockwise until fingertight. DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.