ICS-2100 Ion Chromatography System
90
Doc. 065291-01 3/09
4.5
Pump Does Not Start
No power (front 
Power 
LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace, if needed (see 
No communication between ICS-2100 and Chromeleon
Verify that the USB cable is connected correctly. For connection and setup 
information, refer to the ICS-2100 installation instructions.
Pump is turned off
Turn on the pump from either the Chromeleon Control panel or the ICS-2100 
touch screen 
HOME
 page.
Flow rate is set to 0
Set the flow rate from either the Chromeleon Control panel or the ICS-2100 
touch screen 
HOME
 page.
4.6
No Flow
Pump waste or priming valve open (see 
Close the valves by turning the knobs clockwise until fingertight. DO NOT 
OVERTIGHTEN! Overtightening may damage the valve and the pump 
head.