MS-70 / MX-50 / MF-50 / ML-50 
21 
 
7.2.
7.2.
7.2.
7.2.  Quick Mode Operation
Quick Mode Operation
Quick Mode Operation
Quick Mode Operation    
 
 
  The quick mode can obtain the moisture content with 
settings of ACCURACY and drying temperature. 
 
  Sample is heated up at 200°C for approximately three 
minutes so that moisture content is measured quickly. 
 
7.2.1.
7.2.1.
7.2.1.
7.2.1.  ACCURACY
ACCURACY
ACCURACY
ACCURACY    
 
  ACCURACY of measurement can be set either HI, MID. or LO. 
 
  The sample quantity is automatically selected by ACCURACY. 
 
 The termination value of the analysis mode is automatically selected by 
ACCURACY and minimum scale value of % display. 
 
  Analysis mode is the program to finish the measurement when a change of moisture 
content per one minute becomes smaller than a preset termination value. 
 
  The settings are as follows: Specify an ACCURACY. 
    
ACCURACY 
 Model 
Minimum 
scale
HI 
MID. 
LO 
  0.001 % 
 0.02 %/min
 0.05 %/min   0.05 %/min
  0.01 % 
 0.05 %/min
 0.10 %/min   0.20 %/min
MS-70 
  0.1 % 
 0.10 %/min
 0.20 %/min   0.50 %/min
  0.01 % 
 0.05 %/min
 0.10 %/min   0.20 %/min
MX-50 
  0.1 % 
 0.10 %/min
 0.20 %/min   0.50 %/min
 0.05 
 0.1 
MF-50 
 1 
 0.10 %/min
 0.20 %/min   0.50 %/min
 0.1 
Preset 
Termination 
value 
ML-50 
 1 
 0.20 %/min
 0.50 %/min   1.00 %/min
Sample quantity 
5 g 
2 g 
1 g 
Use Precise 
result 
 Quick 
measurement
 
 
7.2.2.
7.2.2.
7.2.2.
7.2.2.  Operation 
Operation 
Operation 
Operation     
 
  This operation explanation uses the following example of the MX-50 : 
 
  Refer to "8.Selection of Measurement Method" for detail. 
 
  
Input 
Parameters 
 
   
Analysis mode ................................... Quick mode   
 
   
Heating pattern .................................. Quick drying (
, Automatic selection) 
 
   
Drying temperature ............................ 130 °C 
 
   
ACCURACY.......................................  LO  
 
   
Sample quantity ................................. Approximately 1 g (Automatic selection) 
 
   
Analyzing mode ................................. 0.20 %/min (Automatic selection) 
 
 
  Stored Parameters (Factory Settings for the MX-50) 
 
   
Measurement unit ............................... Moisture content is based   
on a wet sample 
 
   
Minimum scale value of % display........ 0.01 % 
 
   
Minimum scale value of gram display ... 0.001 g 
 
   
Data memory function........................ Not used 