A&D MS-70 User Manual

Page of 75
MS-70 / MX-50 / MF-50 / ML-50 
22 
 
 
1.  Turn on the analyzer.   
The gram unit (of weighing mode) is displayed. 
 
 
Enter the Quick Mode
Enter the Quick Mode
Enter the Quick Mode
Enter the Quick Mode    
 2. 
Press 
the 
 SELECT 
 key to display an analysis 
mode and press the 
 or 
 key to select 
 quc 
 
 
 
 
 
Select ACCURACY
Select ACCURACY
Select ACCURACY
Select ACCURACY    
 3. Press 
the 
 SELECT 
  key to select 
ACCURACY.
 
  
When 
ACCURACY 
is selected, HI, MID. or LO blinks. 
 
 4. Press 
the 
 or 
 key to select LO of ACCURACY. 
 
 
 
 
Set Drying Temperature at the Sample Pan
Set Drying Temperature at the Sample Pan
Set Drying Temperature at the Sample Pan
Set Drying Temperature at the Sample Pan    
 5. Press 
the 
 SELECT 
  key to select the drying 
temperature. 
 
 6. Press 
the 
 or 
 key to set 130 °C. 
 
 
 
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation    
 7. Press 
the 
 ENTER 
  key to store the parameters. 
 
  The weighing mode is automatically displayed. 
 
 
 
 
 
 
Put 
Put 
Put 
Put a 
a Sample on the Pan
Sample on the Pan
Sample on the Pan
Sample on the Pan    
  8.  Put the breeze break ring, pan support, pan handle 
and sample pan in order. (With no sample.) 
 
 
9.  Close the heater cover. 
 
  10.  When displaying a stable value, Press the 
 RESET 
 key. 
Avoid mechanical vibration, breeze and environmental 
noise during measurement. If it deviates from zero 
display, press the 
 RESET 
 key. 
 
Close
Preparation 
Quick mode 
ACCURACY 
Temperature 
Quick drying 
ML-50 does not display