  The moisture content is calculated with a ratio of wet weight and dried weight. 
Therefore, the absolute value of weighing does not influence the calculation of the 
moisture content, but it is necessary to get precise weighing for GLP, GMP and ISO. 
Use a 20g mass or a 50g mass to calibrate the weighing sensor. 
 
  When calibrating the weighing sensor, you can output the calibration report adapted 
to GLP, GMP and ISO. 
 
  There is a certified temperature calibrator (accessory AX-MX-43, only for MS-70 and 
MX-50) to calibrate the pan temperature for precise temperature control. 
 
  When calibrating the temperature, you can output the calibration report adapted to 
GLP, GMP and ISO. 
 
  The analyzer can store an ID number to be used in the calibration report. The 
number can be used for management and maintenance of the analyzer 
 
 
 
13.1.
13.1.
13.1.
13.1. Identification Number (ID No.)
Identification Number (ID No.)
Identification Number (ID No.)
Identification Number (ID No.)    
 
 
  The ID number consists of the following seven characters. 
Characters 
4  5
6
7
8
9
Space 
-(hyphen) 
Display 
0  1  2  3  4  5 6 7 8 9
 - 
 
Characters  A B C D E F G H
I
J K L M N O P Q 
Display 
a  b  C  d  e  f g H i j k l m N o p q 
 
Characters R S T U V W X Y Z
Display 
r  s  t  U  v  w x y z
 
 
 
13.1.1.
13.1.1.
13.1.1.
13.1.1.  Setting the ID Number
Setting the ID Number
Setting the ID Number
Setting the ID Number    
 
 
1.  Turn on the analyzer.   
The gram unit (of weighing mode) is displayed. 
 
 
2.  Press and hold the 
 SELECT 
  key to enter the function 
table. Then 
 Cl adj 
 is displayed. 
 
 3. Press 
the 
 SELECT 
  key to display 
 id 
 
 4. Press 
the 
 ENTER 
 key. 
 
 
 
5.  Set the ID number using the following keys.   
Example: lab-123 
 
 
 SELECT 
 key ... Selects a figure. 
 
 
 key............. Selects a value for the figure. 
 
 
 ENTER 
  key..... Stores the ID No. and proceeds to step 6. 
 
 
 6. Press 
the 
 RESET 
  key to return to the weighing mode. 
 lab-123
 
 0.000
g
 
 0.000
g
 
 Cl 
adj
 
 id
 
Press and