13.2.
13.2.
13.2.
13.2. Calib
Calib
Calib
Calibration of the Weighting Sensor
ration of the Weighting Sensor
ration of the Weighting Sensor
ration of the Weighting Sensor    
 
    A standard mass of 20g or 50g can be used for the calibration 
    A 20g standard mass (AX-MX-41) is recommend. 
Caution 
 
  Avoid vibration and drafts that affect the calibration. If affected, the analyzer 
may be unable to calibrate the weighing sensor. 
 
  Use a 20g mass for the calibration, because the height between the weighing 
pan and glass-housing is 26 mm. If a tall mass is used, open the glass-housing 
and avoid external influence. 
 
13.2.1.
13.2.1.
13.2.1.
13.2.1.  Operation
Operation
Operation
Operation    
 
1.  Display the gram unit (of the weighing mode). 
 
 
2.  Install the weighing pan, pan support and pan handle. 
 
  Close the heater cover. 
 
 
3.  Press and hold the 
 PROGRAM 
  key to display 
 CH 
 
 
 4. Press 
the 
 SELECT 
  key to display 
 Cal 
 
 
 5. Press 
the 
 ENTER 
  key to display 
 Cal 0 
 
 
6.  If 20g mass is used, ...... Press the 
 ENTER 
 key. 
Proceed to step 8. 
 
  If 50g mass is used, ...... Press the 
 SELECT 
 key. 
Proceed to step 7. 
 
 7. Press 
the 
 or 
 key to select 50.000g.   
Press the 
 ENTER 
  key to store it.   
 end 
 Cal 0 
  are displayed in order. 
 
 
 
 8. When 
displaying 
 Cal 0 
, press the 
 ENTER 
 key 
to input "Data to display ZERO". The standard 
mass value is displayed (Example: 20g). 
 
 
 
9.  Open the heater cover and put the standard mass 
on the center of the pan and press the 
 ENTER 
   
key to input "mass value".   
 end 
 is displayed. 
 
 
 
 10.  Remove the mass to return to the weighing 
mode 
 
  If the report for GLP, GMP and ISO is to be 
output (Refer to page 60), 
 glp 
 is displayed. 
The output condition for the report is selected in 
the function table. 
 0.000
g
 
 CH
 
 Cal
 
 Cal 
0
 
 50.000
g
 
 end
 
The input data to display ZERO 
The input of mass value 
 end
 
 glp
 
  
20
 
 0.000
g
 
Standard mass 
Preparation 
Remove mass 
Mass value 
Press and hold