A&D MS-70 User Manual

Page of 75
MS-70 / MX-50 / MF-50 / ML-50 
65 
 
15.4.
15.4.
15.4.
15.4. Troubleshooting
Troubleshooting
Troubleshooting
Troubleshooting    
 
1.
In the Case that
In the Case that
In the Case that
In the Case that    PPPProper 
roper 
roper 
roper RRRResults 
esults 
esults 
esults are
are
are
are not be 
 not be 
 not be 
 not be O
O
O
Obtained.
btained.
btained.
btained.    
 
  Use the self-check function. Refer to 9.1.Self-Check Function . 
 
 
  Check repeatability. ( Weigh the same mass several times in 
the weighing mode.) A taller mass may touch the 
glass-housing. Use a short mass if possible. If a tall 50g mass 
is used, open the heater cover and avoid external influence. 
 
 
  The height from sample pan to glass-housing is 26 mm. 
 
 
  Check whether the test sample can be measured correctly. 
 
 
  Avoid the breeze from an air conditioner and vibration. 
 
 
  Check sample condition. Refer to 6.3.Proper Operation for Precision Measurement 
 
 
 Check measurement procedure and pre-heating process. Refer to 6.3.Proper 
Operation for Precision Measurement 
 
 
 
2.
IIIIn Case that 
n Case that 
n Case that 
n Case that the 
the 
the 
the LLLLamp does not 
amp does not 
amp does not 
amp does not LLLLiiiigh
gh
gh
ght or 
t or 
t or 
t or it takes Too Long 
it takes Too Long 
it takes Too Long 
it takes Too Long to 
to 
to 
to RRRReach the 
each the 
each the 
each the DDDDrying 
rying 
rying 
rying TTTTemperature.
emperature.
emperature.
emperature.    
 
  It requires six seconds to light the halogen lamp using 
the 
 START 
 key. 
 
 
  When the heater cover is opened, power is not 
supplied to the halogen lamp. 
 
 
  When an overheat has occurred, power is not supplied to 
the halogen lamp until the halogen lamp becomes cool. 
 
 
  Check the rated voltage of the halogen lamp that is 
printed around the holder. 
 
 
  Read the power supply voltage label on the back of the heater cover and confirm that 
the rated voltage of the halogen lamp is correct for your power supply voltage. 
 
Voltage Label  Power Supply Voltage
The Rated Voltage of 
the Halogen Lamp 
Accessory number
100 - 120 V  AC 100 V to AC 120 V
AC 120 V 
AX-MX-34-120V 
200V - 240 V  AC 200V to AC 240 V
AC 240 V 
AX-MX-34-240V 
 
 
  Is a fuse blown? Check the fuses after removing 
the power cord. Check the rated value and put 
new fuses into the correct holders. 
 
 
  Do you measure a lower drying temperature after 
a high drying temperature? If the lamp is hotter 
than the drying temperature, the measurement 
can not be started. 
 
 
  Check that the sample pan is cool. 
 
 
  Other cases, the halogen lamp may be defective. 
Replace with a new halogen lamp. Refer to "15.2. 
Replacement of the Halogen Lamp". 
 
Maximum height 
of mass is 26 mm.
Read rated 
voltage
Halogen lamp
Fuse, T100mA 250V 
Fuse, T6.3A 250V
Power supply voltage label