16.  Specifications
Specifications
Specifications
Specifications    
 
 MS-70 
MX-50 
MF-50 
ML-50 
Measurement method 
400 W halogen lamp, thermogravimetric analysis 
Drying temperature 
range at sample pan 
50°C to 200°C (1°C increments) 
Heating pattern 
Standard drying, Ramp drying,   
Step drying, Quick drying 
Standard drying,
Quick drying 
Temperature calibration
By Accessory AX-MX-43 
--- 
Sample weight range 
0.1 g to 71 g 
0.1 g to 51 g 
Accuracy: Repeatability of measurement, (Standard deviation) 
 Moisture 
content  1 
   
over 5 g sample 
  0.01 % 
  0.02 % 
  0.05 % 
  0.1 % 
   
over 1 g sample 
  0.05 % 
  0.1 % 
  0.2 % 
  0.5 % 
 Weighing 
mode 
 0.0005 
 0.001 
 0.002 
 0.005 
Minimum reading 
 Moisture 
content 
 0.001 
%, 
 0.01 
%, 
 0.1 
 0.01 
%, 
 0.1 
 0.05%, 
 
 0.1 
%, 
 
 1 
 0.1%, 
 
 1 
 Weighing 
mode 
 0.0001 
 0.001g 
 0.002g 
 0.002g 
Measurement programs 
 
Sample weight and termination value is automatically set with 
ACCURACY and % display. When drying rate reaches the 
termination value, measurement is automatically completed. ( 2) 
 
Standard mode 
Standard drying, Ramp drying, Step drying 
Standard drying
 
Sample weight and termination value is automatically set with 
ACCURACY and % display. When drying rate reaches the 
termination value, measurement is automatically completed. ( 2) 
 
Quick mode 
Quick drying 
 
When drying rate is less than preset termination value, 
measurement is automatically completed. ( 2) 
 
Automatic mode 
Standard drying, Ramp drying, Step drying 
Standard drying
 
After heating for the preset drying time, measurement is 
automatically stopped. (1min. to 480 min.) 
 
Timer mode 
Standard drying, Ramp drying, Step drying 
Standard drying
 
When pressing the key at any time, measurement is stopped and 
the result is decided. 
 
Analys
is
 mode 
Manual mode 
Standard drying, Ramp drying, Step drying 
Standard drying
 
Moisture content (Wet-base) 
 
Moisture content (Dry-base, Atro) 
 Dry 
content 
 Ratio 
 
Measurement unit 
Weight (g) 
  Number of memory 
20 sets 
10 sets 
5 sets 
Data memory function 
100 results 
50 results 
30 results 
Communication