Fluke 566 User Manual

Page of 30
566/568 
Users Manual 
14 
Field of View 
For accurate measurements, make sure that the target is larger than the unit’s spot size. 
The smaller the target, the closer you should be to it (see Figure 6).  
Yes
N
o
 
eyl05.eps 
Figure 6. Field of View 
Emissivity 
Emissivity describes the energy-emitting characteristics of materials. Most organic 
materials and painted or oxidized surfaces have an emissivity of approximately 0.95, the 
default setting for the thermometer.  
To compensate for inaccurate readings that may result from measuring shiny metal 
surfaces, you can cover the surface to be measured with electrical tape or flat black paint 
(<148 
°
C/300 
°
F) with emissivity set to 0.95. Allow time for the tape or paint to reach the 
same temperature as the surface beneath it. Measure the temperature of the tape or 
painted surface. 
If you cannot paint or use tape, then you can improve the accuracy of your measurements 
by either numerically adjusting emissivity or by using the Emissivity Menu to access a 
table of some common materials. The thermometers have numerically adjustable 
emissivity from 0.10 to 1.00 that can be used in conjunction with the emissivities in 
Table 3. Also, the thermometers have a built-in table of the most common materials that 
can be set from the Emissivity Menu. The built-in values are indicated in Table 3. 
HOLD 
The display retains its last infrared measurement for twenty seconds when the trigger is 
released, with HOLD appearing on the display. With the probe inserted, the contact 
thermometer remains on. To freeze the infrared temperature when a probe is not inserted, 
release the trigger until HOLD appears on the display.