Fluke 566 User Manual

Page of 30
566/568 
Users Manual 
16 
External Contact Probe 
XW
Warning 
To avoid electrical shock or personal injury, do not connect the 
external contact probe to live electrical circuits. 
The thermometers come with a bead K-type thermocouple probe. The probes attach to the 
thermometers via the probe input located on the top of the thermometer (see Figure 7). 
With the probe installed, the probe symbol (A) appears on the display. The probe 
can be used simultaneously while the thermometer is taking non-contact measurements. 
The probe readings are shown below the non-contact measurements. The thermometer 
will stay on when there is a probe inserted. When used with a personal computer and 
FlukeView Forms (568 only), the auto off feature is disabled. 
Troubleshooting 
Use the following table to remedy possible problems when using the thermometer. 
Table 4. Troubleshooting 
Symptom Cause  Action 
--- (on display) 
Target temperature is over or 
under range. 
Select target within 
specifications 
Low batteries 
Replace Batteries 
Blank display 
Thermometer is asleep 
Possible dead batteries 
Pull trigger 
Replace batteries  
Laser does not work 
Low or dead batteries 
Ambient temperature is above 
40 
°
C (104 
°
F) 
Replace batteries 
Use in area with lower ambient 
temperature 
Failed communication via USB 
Unit is off  
FlukeView Forms is not running 
Pull the trigger 
Start FlukeView Forms 
Inaccuracy 
Possible incorrect emissivity 
setting, field of view, or spot size 
See “Emissivity”, “Field of View” 
and “Distance and Spot Size” 
sections. 
Settings such as emissivity, 
date/time, F/C, and saved data 
lost 
Battery dead or not replaced in 
<1 minute of removal 
Reset settings. Replace 
batteries as soon as low battery 
indicated; Exchange the 
batteries within one minute of 
removal. Refer to “Changing the 
Batteries”.