Fluke PM6685R User Manual

Page of 276
Command Reference 9-69
:MEASure :TINTerval?
PM6680B PM6681
8 (@«1|2|4»),(@«1|2|4»)]
Time-Interval
Traditional time-interval measurements are performed. The trigger levels are set
automatically, and positive slope is used. The first channel in the channel list is the
start channel, and the second is the stop channel.
Parameters:
The first (@«1|2|4») is the start channel and the second (@«1|2|4») is the stop channel
(@1) means input A
(@2) means input B
(@4) means input E (Rear panel arming input)
If you omit the channel, input A is the start channel, and input B is the stop chan-
nel.