Fluke PM6685R User Manual

Page of 276
9-86 Command Reference
2
Alias commands fot the: SDELay command for compatibility with the PM6680B.
n
Sense Subsystem command tree for PM6685
[:SENSe]
:FUNCtion
8 ‘Measuring function [_ Primary channel [ , Secondary channel ] ] ‘
:EVENt
1
:LEVel
8 <Trigger level in Volts> | MIN | MAX
:AUTO
8 ON | OFF | ONCE
:HYSTeresis
8 <Sensitivity band in Volts> | MIN | MAX
:AUTO
8 ON | OFF | ONCE
:SLOPe
8 POS | NEG
:ACQuisition
:APERture
8 <Measurement Time> | MIN | MAX
:HOFF
2
:TIME
8 <Hold off time> | MIN | MAX
:AVERage
:STATe
8 ON | OFF
:ROSCillator
:SOURce
8 INTernal | EXTernal
:SDELay
8 <Burst sync. delay> | MIN | MAX
:TOTalize
:GATE
[:STATe]
8 <ON | OFF
:INTernal
:FORMat
8 REAL | PACKed
1
Alias commands for commands in the Input subsystem.