Fluke PM6685R User Manual

Page of 276
9-88 Command Reference
:ACQuisition :HOFF
PM6680B PM6681
8 <boolean>
Hold Off On/Off
Switches the Hold Off function On/Off.
Parameters:
<Boolean> = 1 / ON | 0 / OFF
Returned format:
1 | 0¿
*RST condition:
OFF
:ACQuisition :HOFF: ECOunt
PM6680B PM6681
8 «<Decimal value>|MIN|MAX»
Hold Off, set event counter
Sets the Hold Off event value. The counter counts negative events on the B input
(channel 2).
Parameters:
<decimal value> is a number in the range 2 to 16 777 215.
Returned format:
<Decimal value>¿
*RST condition:
100