Standard programs
5.6.2 Take a color difference measurement with stored reference values
1.
Press Color Difference Measurement from the Main Menu.
2.
Press Select Reference.
3.
A list of available references is displayed.
4.
Press the required reference.
5.
Press Select Reference to confirm.
6.
Insert the test cuvette.
Important note: For the color difference measurement, it is
absolutely necessary that the type of cuvette/sample cell used for
the sample measurement is identical to the type of
cuvette/sample cell that was used for the reference
measurement.
absolutely necessary that the type of cuvette/sample cell used for
the sample measurement is identical to the type of
cuvette/sample cell that was used for the reference
measurement.
Example:
It is not possible to select a reference that was measured in a
10 mm cuvette, and then run a difference measurement with a
50 mm cuvette.
10 mm cuvette, and then run a difference measurement with a
50 mm cuvette.
5.6.3 Add a reference to the reference list
1.
Press Color Difference Measurement from the Main Menu.
2.
Insert a reference cuvette/cell.
3.
Press Options >More > Store Reference. The Reference
Data list is displayed.
Data list is displayed.
4.
Press a free reference number.
5.
Press Store.