Symmetricom 5401 User Manual

Page of 40
PN 8500-0150, Rev. A, May 2005
23
Operating Manual
Operation
The srr.rrrr is a number from -99.9999 to +99.9999 microseconds per second. The number is 
frequency change from DAC = 00000 to 65535 of the unit. This number is used in the 
discipline algorithm to help compensate for oscillators with different frequency 
Response Format: 
Design Verify On/Off 
Command Format: 
Description: 
Determines whether filtered rate or filter offset is returned in a status request. When x = 0, the 
extra data is not included. When x = 1, the extra data is included. 
The response to the status command is changed to the following: 
Frr.rrrrrr - Filtered rate in microseconds; rr.rrrrrr is the an average frequency rate error 
calculated internally by the unit. This is one of the values used by the program to determine 
the dynamic gain. 
Oyyyyyy.yyy - Filtered offset in microseconds; yyyyyy.yyy is the an average time error 
calculated internally by the unit. This is one of the values used by the program to determine 
the dynamic gain. 
Response Format: 
Note: It is assumed that the DAC output and oscillator control are linear. If not, then the 
oscillator control range should be extrapolated from a subset of the control curve that is 
near zero frequency error.