Adjustment Procedures
5-10
AWG710 Service Manual
This procedure uses the general purpose knob on the front panel to set the Magic
Frequency parameter.
Equipment
Required
One oscilloscope (Item 1)
One SMA 50 W coaxial cable (Item 4)
SMA (Fe) to BNC (Ma) 50 W adapter (Item 5)
Do the following steps to connect the test equipment and set test equipment
controls:
1. Use a SMA coaxial cable and a SMA (female) to BNC (male) adapter to
connect the waveform generator CH1 Output connector to the CH1 input
connector on the oscilloscope.
SMA 50 W coaxial cable
AWG710 Arbitrary Waveform Generator
Oscilloscope (TDS784D)
ÁÁ
ÁÁ
CH1 input
CH1 output
SMA(Fe)ĆBNC(Ma)
adapter
Figure 5-4: Hookup for the marker level adjustment
2. Set the Oscilloscope controls as follows:
Vertic al . . . . . . . . . . . . . . . . . . . . . . . . . CH 1
CH 1 coupling . . . . . . . . . . . . . . . . DC
CH 1 sc ale . . . . . . . . . . . . . . . . . . 200 mV/div
CH 1 input impedance . . . . . . . . . . 50 W
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . 200 ns/div
Trigger
Source . . . . . . . . . . . . . . . . . . . . . CH 1
Coupling . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . Auto
Magic Frequency