Tektronix 070-A830-51 User Manual

Page of 267
Adjustment Procedures
AWG710 Service Manual
3. Do the following to set the waveform generator controls:
Press UTILITY 
 System  Factory Reset  OK.
4. Follow the substeps below to load the waveform:
a. Press SETUP 
 Waveform/Sequence  Load...
b. Turn the general purpose knob to display the list of waveform files and
highlight the file MODE.WFM.
Press OK to select and load the file.
c.
Press the RUN button.
d. Press the CH1 OUTPUT button to ON.
5. Adjust Magic Frequency:
a. Press UTILITY 
 Service
b. Turn the general purpose knob to select Tweak AWG2.
c.
Press OK.
d. Press Magic Freq. .
e.
Turn the general purpose knob to adjust the waveform on the oscillo-
scope display so that the corruption of the waveform becomes the
maximum.
6. Press Save to save the setting.
7. Disconnect the oscilloscope.