FMA 3400/3400ST Series Dip Switches 
 
 
 
 
 
Dip switch 4 is allocated to the primary calibration gas. 
Dip switch 5 is allocated to the second calibration gas (if applicable). 
Dip switch 6 is allocated to the third calibration gas (if applicable).  
 
To select the gas, the dip switch should be turned ON. All other switches 
allocated to gases (i.e. 4, 5 or 6 except the required switch/gas) should be 
set to OFF. 
 
 
D.
 
Maintenance and Product Care 
 
1.  General  
 
Inlet filters should be periodically checked and cleaned or replaced as 
necessary. 
 
Regularly check all electrical and process connections for damage or 
deterioration. 
 
If the sensor is to be stored, keep both the inlet and outlet ports sealed. 
 
Do not allow any liquid or moisture to enter the sensor or damage will 
occur. 
 
 
2.  Returning Units for Repair or Recalibration 
 
To return a unit for repair or recalibration please contact the Omega 
Engineering Customer Service Department. An Authorized Return (AR) 
number will then be issued. The AR number should then be noted on the 
outside of the package and on any correspondence. Further details may be 
found on page 25 of this manual.