If a gas other than the calibration gas is used then the adjusted maximum 
(full scale) flow for the unit should be calculated using the K Factor for 
that gas (see section C3 above).   
 
A zero or negative set-point voltage will cause the solenoid valve to close 
fully. Whilst closed, the valve is configured to withstand pressures up to 60 
psig (higher pressures on request).  
 
 
Caution: The flow controller valve will open if the pressure 
exceeds 60psig. For safety it is recommended that a separate 
positive shut-off valve is installed upstream of the controller. 
 
 
 
5.  Changing The Flow Rate Set-Point – FMA3400/3400ST Series 
Only 
 
On the FMA 3400/3400ST Series the set-point may be input from an 
external source or be supplied internally.  
 
For an external set-point, dip switch 1 should be OFF and dip switch 2 
ON. See section C4 above for details of how to adjust the set-point using 
an external voltage source. 
 
For an internal set-point, dip switch 1 should be ON and dipswitch 2 OFF
Adjustment of the internal set-point is made by turning the coarse and fine 
set-point potentiometers on the front panel of the display, with the gas is 
flowing, until the desired flow rate is achieved.  
 
FMA 3400/3400ST Series Set-Point Potentiometers 
 
 
 
 
6.  Power Save Mode. 
 
To improve valve performance and reliability over time, the FMA 
3200/3400 Series features a Power Save Mode. This is activated after a 
prolonged application of a zero or negative set-point. When a control 
voltage greater than 0 VDC is applied after the Power Save Mode has been 
initiated there may be a short delay (1-2 secs) before the valve actuates.