RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 112 
To access Q-RPT calibration information viewing or editing press [SPECIAL],  <8Cal>.  
Select the desired Q-RPT.  The <1view> selection displays the calibration information fields.  
The <2edit> function displays the fields and allows them to be edited.  The display is: 
1. 
Date of last calibration in YYYYMMDD format. 
2. 
Position designator of the Q-RPT being viewed. 
3. 
Current value of absolute mode P
offset
                        
         
 
Cal date:20050425 Hi 
AbsPoffset: 2410 Pa 
                           
 
 
If in edit mode, the calibration information fields can be edited.  Edits to P
offset
 are common 
with the AutoZ edit or run function (see Sections 3.3.9, 3.5.1).  Pressing [ENT] on the last 
field goes to the next view/edit screen: 
1. 
Value of PA. 
2. 
Position designator of the Q-RPT being viewed. 
3. 
Value of PM. 
           
                      
 
PA: 0.0 Pa        Hi 
PM: 1.000000 
                 
 
 
If in edit mode, the calibration fields can be edited.  If the selected Q-RPT is an Axxx Q-RPT, 
pressing [ENT] in the PM field goes to the next calibration information display (this display is 
skipped if the Q-RPT is a Gxxx or BGxxx type): 
1. 
Position designator of the Q-RPT being viewed. 
2. 
Cursor is on current ON or OFF selection. 
                                    
 
Allow absolute    Hi 
meas mode:  1on 2off 
                          
 
 
The cursor is on the current setting.  If in edit mode, the ON or OFF status can be changed 
by selecting <1on> or <2off> (see Section 5.3.5).  [ENT] in view mode returns to the view/edit 
screen.  [ENT] in edit mode goes to confirmation of change activation if changes have been made 
in any of the calibration screens.   Pressing [ESC] in any edit screen exits the edit screen without 
activating any changes. 
 
  The value of PA is always in Pascal (Pa).  The value of PM is dimensionless. 
 
5.3.8 
Q-RPT ADJUSTMENT WITHOUT CALTOOL FOR Q-RPTS 
SOFTWARE 
 PRINCIPLE 
The RPM4 Q-RPTs can be adjusted without using CalTool for Q-RPTs software.  This requires: 
• Setting 
P
offset
 to zero. 
•  Applying pressures with a calibration standard and recording the pressures measured by RPM4. 
•  Calculating new PA and PM values and entering them.