Fluke A70M User Manual

Page of 148
 
5.  MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS 
 
Page 113 
© 2007 DH Instruments, a Fluke Company 
  Before proceeding to adjust a Q-RPT without using CalTool for Q-RPTs software, Sections 
5.3.1, 5.3.2, 5.3.3 and 5.3.4. 
 
 OPERATION 
The typical procedure for adjusting a Q-RPT is: 
  Set-up and prepare the RPM4 for calibration (see Sections 5.3.2, 5.3.3).  
 Use 
the 
[RANGE] function key to select the Q-RPT to be calibrated (see Section 3.3.1).  
Be sure to set the DF range (not an AutoRanged range). 
 Use 
[HEAD] to set the HEAD to zero (see Section 3.3.7).  Use [UNIT] to set the desired 
pressure unit of measure (see Section 3.3.2).  Use [MODE] to set the desired pressure 
measurement mode (see Section 3.3.3). Q-RPTs that will be used in gauge 
measurement mode only after the adjustments are made, may be calibrated in gauge 
measurement mode with a gauge pressure standard (see Section 5.3.5). 
 Use 
[SPECIAL],  <1AutoZ> to access the AutoZ function (see Section 3.5.1).  Turn 
AutoZ OFF.  If you are only verifying the Q-RPT, “as received” prior to adjustments, leave 
AutoZ ON. 
 Use 
[SPECIAL]<8cal><1Hi Q-RPT > or <2Lo Q-RPT ><1edit>, to read and record the 
current values of PA and PM for the Q-RPT to be calibrated.  If verifying a calibration in 
absolute mode, also record the value of P
offset
  Run the recommended calibration point sequence for the Q-RPT (see Section 5.3.4). 
Record the pressure applied by the standard and the RPM4 reading at each calibration point.  
Dwell at least 60 seconds at each point after setting the reference pressure to allow full 
stabilization. 
  Enter the calibration pressures and RPM4 readings into a spreadsheet.  Calculate the 
“non-corrected” RPM4 readings by backing out the PA and PM recorded in Steps   and   
above, following: 
non-corrected reading = (corrected reading - PA)/PM 
If calibrating in absolute mode, AutoZ was ON and the value of P
offset
 was not set to zero, 
also add the value of P
offset
 to each reading. 
  Perform a linear regression to find the offset and slope that best fit the non-corrected 
RPM4 readings to the calibration standard pressures.  The offset is the new value of PA, 
the slope is the new value of PM. 
 Press 
[SPECIAL]<8Cal><1Hi Q-RPT > or <2Lo Q-RPT ><2edit> and write the new 
calibration date and the new values of PA and PM for the Q-RPT calibrated (see Section 
5.3.7).  If the calibration was performed in gauge mode, turn absolute mode OFF so it won’t 
be used after the calibration. 
  Calculate as left data for the calibration if desired following: 
as left reading = (non-corrected reading • new PM) + new PA 
  Perform additional verification pressure runs as desired.