 PRINCIPLE 
RPM4 A70M/A20M-AF is designed to support the calibration and test of a very wide variety of 
test ranges in absolute and gauge measurement modes. 
The AutoRange function simplifies the task of selecting the Q-RPT best suited to cover a 
specific test range and setting measurement parameters appropriately for that specific range.  
These are set automatically based on operator entry of desired measurement mode, pressure unit 
of measure and range full scale.  The selections and settings made by AutoRange are summarized 
in Table 4. 
A range set up using AutoRange and all its range specific settings, may be saved for recall 
using [SETUP]<1range> (see Section 3.4.1).  Saved ranges are available for later selection 
using  [RANGE]  (see Section 3.3.1).  If a range created by AutoRange is not saved, it is 
overwritten by the next AutoRange or erased when another range is selected. 
 
  Press [ENT] from the main run screen in Rate DISPLAY mode to use the [AutoTest] 
function to AutoRange RPM4 based on DUT tolerance (see Section 3.3.10). 
 
Table 4.  Settings made by AutoRange 
SETTING/ 
SELECTION 
PURPOSE 
SETTING BY 
[AUTORANGE] 
SETTING BY 
[AUTOTEST[ 
(see Section 3.3.10) 
RELATED 
SECTION 
Unit 
Set pressure unit of 
measure 
Operator specified 
during AutoRange 
Operator specified in 
AutoTest definition 
3.3.2 
Mode 
Set pressure 
measurement mode 
(absolute, gauge) 
Operator specified 
during AutoRange 
Operator specified in 
AutoTest definition 
3.3.3 
Full scale 
Establish span 
maximum 
Operator specified 
during AutoRange 
Maximum pressure of 
AutoTest definition 
None 
Q-RPT 
Select most 
appropriate Q-RPT to 
cover range and mode. 
Q-PRT with lowest 
default full scale that is 
greater than the 
AutoRange full scale 
Q-PRT with lowest 
default full scale that is 
greater than the 
maximum pressure of 
AutoTest definition 
1.2.2.1 
Upper limit 
Set maximum pressure 
alarm limit. 
105 % of AutoRange 
full scale pressure 
105% of maximum 
pressure of AutoTest 
definition 
3.4.4 
Resolution 
Set display resolution 
10 ppm of AutoRange 
span or  
1 ppm of default Q-
RPT span, whichever is 
larger 
10% of DUT tolerance 
or 
1 ppm of default Q-
RPT span, whichever is 
larger. 
3.4.2 
Stability Limit 
Set Ready/Not Ready 
stability criterion. 
50 ppm of AutoRange 
span or 2 ppm of 
default Q-RPT span, 
whichever is larger 
10% of DUT tolerance 
or 
2 ppm of default Q-
RPT span, whichever is 
larger 
3.4.3, 
3.2.1