3. OPERATION 
 
Page 29 
© 2007 DH Instruments, a Fluke Company 
  OPERATION 
To use the AutoRange function, press [AutoRange] and respond to the RPM4’s prompts. 
  Select AutoRange measurement mode: The screen is identical in appearance and 
function to the [MODE] screen (see Section 3.3.3). 
  Select pressure unit of measure:   The screen is identical in appearance and function 
to the [UNIT] screen (see Section 3.3.2). 
  Enter the AutoRange full scale pressure: Enter the full scale pressure value in the  
<------> entry field and press [ENT].  If the full scale value entered cannot be covered by 
RPM4, an error message appears.  If the full scale value can be covered, AutoRange 
proceeds to the range display. 
  View proposed AutoRange range:  The proposed AutoRange range screen is: 
1. 
Indication that this is a screen of the proposed AutoRange range. 
2. 
Designator and position of the Q-RPT that AutoRange has selected 
to cover the AutoRange full scale and measurement mode. 
3. 
 Pressure unit of measure specified in AutoRange. 
4. 
AutoRange full scale pressure in gauge mode and absolute.  
      
                           
 
AutoRange  A70M   Hi
psi  5000g/5000a 
  
              
             
 
 
  If AutoRange selects a Q-RPT but you would like to use the other one, the [
]
 and [
cursor 
control keys may be used to toggle between Q-RPTs so that the other can be selected. 
  Accept proposed AutoRange:  To accept the proposed AutoRange and go to the MAIN 
RUN screen with the new AutoRange range active, press [ENT] or [AutoRange].  To 
modify the AutoRange full scale, measurement mode or pressure unit of measure, use 
[ESC] to back through the AutoRange screens and make changes. 
 
 
  The HPMS Lo Q-RPT Isolation Valve protects the Lo Q-RPT (3 000 psi max) from 
overpressure when the Hi Q-RPT is in use.  Familiarize yourself with HPMS operation (see Section 
3.2.5) before changing ranges and applying pressures.  Always turn the Lo Q-RPT Isolation Valve 
fully in the direction of the lighted green LED before applying pressure. 
 
  To verify or confirm the range from the MAIN RUN screen, use [+/-] or [RANGE] to view the 
active range (see Section 3.3.1). 
 
  See Table 4 for a listing of all selections and settings affected by AutoRange and their values. 
3.3.5 
[LEAK CK] 
 PURPOSE 
To run an automated leak check routine that determines the leak rate of the system 
connected to the RPM4/HPMS TEST port. 
 PRINCIPLE 
The LEAK CHECK function is provided as a means of checking and quantifying the leaks that 
may be present in the system connected to the RPM4/HPMS TEST port.