The display is: 
0.0 psi g         Lo 
Exercise DUT & [ENT]  
 
The top line is the standard RPM4 run screen indicating the current measured 
pressure (see Section 3.1.1). Exercise the DUT if desired using the top line indication 
of the measured pressure and press [ENT] when complete.The display is: 
1. 
Standard RPM4 run screen indicating current 
measured pressure. 
2. 
Number of this test point/the total number of 
test points in the sequence. 
3.  Pressure to set for this increment in current 
pressure unit of measure and measurement 
mode. 
       
   
0.0 psi g         Lo 
Set 0.0         1/11 
         
                      
 
 
Adjust the pressure to the <Set nnnn> value indicated.  Set the pressure so 
that the DUT, NOT the RPM4, indicates the <Set nnnn> value.
 
Once the specified pressure has been set, press [ENT].  If the RPM4 pressure 
reading is in Not Ready condition (see Section 3.2.1), the RPM4 makes an 
invalid entry sound and displays <Pressure NOT READY, TRY AGAIN>.  
Correct the conditions that are causing the RPM4 reading to be excessively 
unstable and press [ENT] again.  If the RPM4 pressure reading is in Ready 
condition, the RPM4 makes a valid entry sound and the display is: 
1. 
Standard RPM4 run screen indicating current 
measured pressure. 
2. 
Number of this test point/the total number of 
test points in the sequence. 
3. 
<F> to indicate freeze (see Section 3.3.6.7) 
followed by the RPM4 pressure reading in 
current units of measure captured when [ENT] 
was pressed in the previous <Set> screen.  
This is the reference pressure that was 
applied to the DUT when the DUT indicated 
the cardinal point.  The frozen pressure value 
flashes if the reading was out of tolerance 
(definition of out of tolerance is:  Set point – 
RPM4 reading > DUT tolerance).. 
       
   
497.7 psi g       Lo 
F497.6    OT   2/11 
      
           
         
 
 
Press [ENT] to accept this test point and proceed to the next test point or [
] to 
repeat the point. 
Repeat the set pressure, take reading and accept data procedure until all points 
in the sequence have been accepted. 
After the last pressure point in the sequence 
has been accepted the display is: 
AutoTest complete 
1data 2new 3repeat 
 
Select  <1data> to view the data collected in the AutoTest that was just 
completed (see Section 3.4.5.1). 
Select <2new> to run a new Quick or File AutoTest. 
Select <3repeat> to repeat the AutoTest that was just completed. 
Press [ESCAPE] to return to the main run screen.